Connect power plug to the socket. Put ON the Mains on the control panel, temperature controller displays and the inside chamber motor ON. Set the thermostat at 5°C above the Set Temperature on the temperature controller. The Safety thermostat will cut off the mains contactor in case of overshoot of the temperature by 5°C.
Set the Print time interval for 30 minutes on PC /Printer interface unit and connect the same to 80 column printer or to the PC for data communication. Put ON the Heating Switch on the Control Panel, Dry heater will be ON.
Put ON the Cooling Switch on the Control Panel, Refrigeration condenser motor and compressor will be ON. Enter the temperature into Temperature Monitoring record as per Annexure – I. The temperature should not vary? 1°C against the set temperature
Procedure, Operation, Calibration, BOD Incubator, Microbiology
Set the Print time interval for 30 minutes on PC /Printer interface unit and connect the same to 80 column printer or to the PC for data communication. Put ON the Heating Switch on the Control Panel, Dry heater will be ON.
Put ON the Cooling Switch on the Control Panel, Refrigeration condenser motor and compressor will be ON. Enter the temperature into Temperature Monitoring record as per Annexure – I. The temperature should not vary? 1°C against the set temperature
Procedure, Operation, Calibration, BOD Incubator, Microbiology