Operation Cleaning and Calibration of Double Door Autoclave

Check the water level before starting the autoclave. If the level is down, then add purified water up to the level mark. Open the door at non-sterile area and load the necessary items to be sterilized as per Appendix – I / appendix –II as required.
Place the autoclave indicator strips at different places in the autoclave for each cycle to indicate the proper sterilization by color differentiation of the strip. Close the door using the radial locking handles clockwise till it is tight. Set the thermograph recorder. Switch on the system to record the temperature cum time of each cycle.
Switch on the mains to supply t
he power followed by the toggle switch located in the control panel to activate the heat generator. Set the multi port-operating valve in slow exhaust position and allow the steam to build in the outer jacket till the pressure reaches 14-17 PSI.
Operation, Cleaning, Calibration, Double Door Autoclave, Microbiology

ichapp

Comments

Popular posts from this blog

OFLOTAL - OR

Ofloxacin - 200mg + Orinidazole-500mg, Model of pack:10x10's, Packing:BlisterOfloxacin - 200mg + Orinidazole-500mg, OFLOTAL - OR TABLETS

Procedure for Inward Material Entries at Security

As soon as a Raw / Packing material consignment arrives at the factory gate, security personnel shall confirm that the material Security shall confirm the place of delivery from the concerned warehouse. Security shall send the documents to the concerned warehouse. Upon receipt of the documents warehouse shall check and ensure the following: Material is for Purchase order exists for the supply. Whether the vendor is approved vendor or not. Certificate of Analysis is enclosed. Invoice / Delivery challan is enclosed. Packing list, if any Lorry Receipt, if any Relevant Excise documents, if any In case of any discrepancy, warehouse shall contact purchase department to resolve the issue. Upon resolving the issue, warehouse shall instruct the security department to allow the material to allow the material inside. Security shall enter the details in the concerned area material inward register as per format no.  xxxxx Security shall put an inward stamp on the back side of the Invoice /...

CARBORIN Z

Carbonyl Iron-100mg+FolicAcid-0.5mg+B12-mcg+Zinc 61.8mg, Model of pack:5x3x10's, Packing:Blister Carbonyl Iron-100mg +FolicAcid-0.5mg+B12-mcg+Zinc 61.8mg, CARBORIN - Z CAPSULES