Connect the power plug to the socket. Check the manometer reading initially that oil level at zero. Put on the power switch to start up. Two switches are present at right below of the instrument. Switch (1 - Red) for air flow.
Switch (2 - Orange) for UV light. Switch (3 - Green) for tube light. Put ON the switch (1 -Red) to start the airflow and manometer will show the reading of Airflow.
Manometer reading should not be less than 10mm water gauge scale while laminar airflow is running. Enter the details of manometer reading in Annexure -I Put ON switch (3 -Green) to start normal light. Put ON switch (2 -Orange) to start UV light.
Operation, Cleaning, Calibration, Garment Cabinet, Microbiology
Switch (2 - Orange) for UV light. Switch (3 - Green) for tube light. Put ON the switch (1 -Red) to start the airflow and manometer will show the reading of Airflow.
Manometer reading should not be less than 10mm water gauge scale while laminar airflow is running. Enter the details of manometer reading in Annexure -I Put ON switch (3 -Green) to start normal light. Put ON switch (2 -Orange) to start UV light.
Operation, Cleaning, Calibration, Garment Cabinet, Microbiology