Operation And Calibration Of (Calibration Of Polarimeter RUDOLPH Autopol V)

Switch ‘ON’ the instrument, from the switch located on the power entry module. Wait till the reading is stable then press “Zero” key from keypad Select the wavelength by pressing “λ” key from keypad and press “Zero” key.
Wash the Temp Troll sample tube with the solvent, which is used in the sample preparation and place in the sample compartment after filling with the same solvent by syringe, as far right as possible in the trough.
Polarimeter, Autopol V, soft cloth, stabilizing temperature, Transmission surface